Abstract

Over 300 daily environmental radioxenon samples were analyzed using French developed SPALAX for automatic sample preparation including high-resolution γ-spectrometry. The 133Xe sensitivity was significantly better than 1 mBq/m 3 (specified criterion for Comprehensive Nuclear-Test-Ban Treaty verification). Radioxenon analysis was extended to include the X-ray region by improved detector window, sample cell design, efficiency calibration, line shape fitting and background analysis. The resulting analysis offers a 4–16 fold improvement in sensitivity for 133mXe and 131mXe, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.