Abstract
The detection sensitivity of various size germanium detectors to low-level neutron dose equivalent rate is described. Six high-purity germanium detectors with a crystal volume of 60 to 120 cm 3 were irradiated by neutrons produced from a 0.76 μg 252Cf isotopic neutron source. The counting rate of the prompt gamma rays of 596 keV was used to determine the detecting sensitivities of the detectors. The results show that the detecting sensitivity of the HPGe detector decreases with increasing detector size, and approximately proportional to the crystal volume.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.