Abstract

A capillary gas chromatographic method using mass spectrometric detection was developed and validated for the trace analysis (ppm level) of methyl methane sulfonate and ethyl methane sulfonate in pharmaceutical drug substances. The method utilizes a capillary column (DB‐624) with 6% cyanopropyl phenyl and 94% dimethyl polysiloxane stationary phase. A dissolve‐and‐injection approach was adopted for sample introduction in a split less mode. Mixture of (80:20) ratio of methanol and chloroform was used as a diluent or sample solvent. A limit of detection of about 0.17 μg/g (0.17 ppm) for methyl methane sulfonate and 0.18 μg/g (0.18 ppm) for ethyl methane sulfonate were achieved and limit of quantitation of 0.52 μg/g (0.52 ppm) for methyl methane sulfonate and 0.54 μg/g (0.54 ppm) for ethyl methane sulfonate were achieved for alkyl sulfonates in drug substance samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.