Abstract
Determination of low concentrations of Ga in environmental or biological samples can benefit from interference free measurement of both Ga isotopes. Unfortunately, both isotopes have potential interferences, i.e doubly charged Ba at m/z 69 and MnO at 71. Analysis using collision and reaction gases by ICP-QQQ as an alternative to HR-ICP-MS is investigated here using conventional nebulization and a desolvating nebulizer. Analysis at m/z 71 is not appreciably affected by MnO at 200 ug/l Mn or in either water reference material; excellent detection limits of < 0.1 ng/l were obtained for He, H2, or on-mass O2 and NH3 gas modes. Analysis of Ga at m/z 69 was severely affected by doubly charged Ba across all gas modes except +17 mass-shift (i.e. 69 - 86) NH3 gas mode. Despite Ga being only 1-2% reactive at the 69-86 mass shift, the doubly-charged interference was not mass-shifted and detection limits of < 1 ng/l and precise and accurate quantification of two water reference materials and nine natural water samples were achievable for this analysis mode.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.