Abstract

The variable background is an important disturbing factor for low-level activity measurements by beta counting. In order to overcome this difficulty in a simple manner a software-aided technique of simultaneous counting in anticoincidence and difference modes has been established. The technique is based on a programmable multi-channel scaler provided with appropriate software for data analysis. Using this counting system, the technique has been tested by performing low-level activity measurements both in a stable and a time-dependent background.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.