Abstract

Proton radiography is one of the most important and actual areas of research that can significantly improve the quality and accuracy of proton therapy. Currently, the calculation of the proton range in patients receiving proton therapy is based on the conversion of Hounsfield CT units of the patient's tissues into the relative stopping power of protons. Proton radiography is able to reduce these uncertainties by directly measuring proton stopping power. The study demonstrates the possibility of Protom synchrotron-based proton therapy facilities to operate in a special mode which makes it possible to implement proton radiography. This work presents the status of the new low beam intensity extraction mode. The paper describes algorithms of low flux beam control, calibration procedures and experimental measurements. Measurements and calibration procedures were performed with certified Protom Faraday Cup, PTW Bragg Peak Chamber and specially designed experimental external.