Abstract

A high surface area, Pt-Ir coating was electrodeposited onto the contacts of Advanced Bionics HiFocus 1j electrode arrays. The coating was imaged using optical and scanning electron microscopy (SEM) and the impedance was directly measured using clinically relevant voltage transients of biphasic pulses as short as 17 μs. The coating, which SEM showed to completely cover the underlying substrate, decreased the polarization impedance of 17 μs and 32 μs pulses by 91% and 93%, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.