Abstract

The measurement system consists of three main components. The magnetic sensor is realized by using a superconducting quantum interference device. To apply all the required magnetic fields and field gradients a coil set-up is placed inclusive current sources. And, a PXI based data acquisition and timing completes the set-up. Each measurement starts with a pre-polarizing of sample volume by applying up to 50 mT. To get the spatial information phase and frequency encoding is used. After the polarization the phase gradient is applied for adjustable time. Then the applied frequency gradient and detection field are inversed repetitively according to the phase time. So, echo signals of samples T2 decrease can be measured.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.