Abstract

Spinal deformities affect young individuals predominantly girls who are usually subjected to regular and intensive radiological investigation especially before and after corrective surgery. Optimisation of spine computed tomography (CT) and the implementation of the low-dose CT in the work-up of spinal deformities were presented. The presented low-dose CT here means providing the operating surgeons with essential information about 15 vertebral bodies (almost 36-cm long region of the vertebral column). The mean effective dose of the low-dose CT was 0.37 mSv without any negative impact on image quality with regard to answering the clinical questions at issue. Tube current modulation (angular and longitudinal) has contributed to 19 % of the total dose reduction and soft tissue algorithm has helped to reduce the artefacts from the metal implants in the postoperative CTs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.