Abstract

The objective of this study was to evaluate image quality of low dose in noncontrast paranasal sinus computed tomography (CT) using single volumetric 320-row multidetector CT technique. The low-dose protocol including tube voltage of 135 kV and tube current of 5 mAs was chosen based on results of the present phantom study. Forty-six patients were assigned to control group with factory standard settings (120 kVp, 75 mAs), and 46 patients were assigned to study group and underwent noncontrast CT of paranasal sinus with low-dose protocol using single volumetric 320-row multidetector CT device. Objective and subjective image qualities were performed. Effective radiation dose calculated for control group scans was 0.45 (SD, 0.39) mSv. It was 0.038 (SD, 0.004) mSv for study group scans. The effective radiation dose of study group was statistically significant lower than control group (P < 0.001). Noncontrast paranasal sinus CT imaging can be performed at very low radiation exposure maintaining high image quality with 135 kVp and 5 mAs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.