Abstract

X-ray computed tomography (CT) with iodine contrast agent is widely employed to locate cancers. However, this method has shortcomings such as high-radiation dose exposure, iodine side effects, and a beam hardening effect. We have been working on the energy-resolved CT measurement method using a novel X-ray detection system, the “transXend” detector, which measures X-rays as electric currents and gives the energy distribution of incident X-rays after analysis. In the present study, we propose a method for low-dose exposure CT that involves the combination of the energy-resolved CT method, which is free from the beam hardening effect, and a harmless contrast agent with high-energy K-edge absorption, such as gold nanoparticles expected as a future contrast agent. Comparisons of radiation dose exposures as functions of aluminum filter thickness at the exit aperture of an X-ray tube and the K-edge energies of contrast agents are described.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.