Abstract

The aim of this study was to investigate whether a low-dose cone-beam computed tomography (CBCT) protocol provides diagnostically acceptable image quality for assessing bone healing after alveolar bone grafting. The study cohort comprised 11 patients (aged 7-14 years) with orofacial clefts who had undergone alveolar bone grafting at Skåne University Hospital in Malmö, Sweden. During the postsurgical follow-up at 6 months, each patient was assessed twice: once with a standard-dose CBCT protocol and once with a low-dose CBCT protocol, which in total corresponds to one CBCT examination made with the exposure settings recommended by the manufacturer. Among others, the assessed parameters included subjective image quality, as well as bone graft height, thickness, and integration. No significant differences were found between the standard- and low-dose protocols for most parameters (p > 0.05). Exceptions included subjective image quality (one observer, p = 0.05) and confidence levels during the assessment (three observers, p = 0.01, 0.01, 0.02). The low-dose protocol yielded adequate image quality for postoperative CBCT healing assessment in patients who have undergone alveolar bone grafting. However, the confidence level of observers during the assessment with the low-dose protocol was reduced. This study is registered on ClinicalTrials.gov (NCT06395077). This study is registered on ClinicalTrials.gov (NCT06395077).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.