Abstract

The aim of the study was to compare radiation exposure and image quality between dedicated computed tomography (CT) protocols for preoperative total hip arthroplasty (THA) planning. Three protocols with automated tube current modulation using 64-slice (n = 177) and 128-slice CT scanners without (n = 129) and with automated tube voltage preselection (n = 84) were compared. All 390 CTs were of sufficient quality for THA planning. Mean DLP was 235.0 mGy*cm (effective dose 2.8 mSv). Lowest radiation exposure (2.5 mSv) was seen with automated voltage preselection and the algorithm's selection was 100 kV (90.5% of patients) and 120 kV. Lowest image noise was seen in the highest dose group (3.1 mSv, 128-slice CT fixed tube voltage). A significant difference in cortical bone radiodensity was seen between 100 kV and 120 kV (P < 0.0001). Preoperative pelvic CT for THA planning is possible with very low radiation dose and reliable quality. Automated voltage preselection further decreases the effective dose by 18.2%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.