Abstract

We propose a new, low-cost pulsatile ventricular assist device (VAD) for short-term applications. The new device could prove very useful in emergency ventricular failure in which patient survival is not assured. In these cases, the device allows ventricular function to be maintained as the patient's situation is evaluated and a decision is made on whether to perform a heart transplant or to replace the device with a long-term VAD. The device has a pneumatic tubular blood chamber, clip valves over the cannulae, and a compliant input chamber that improves filling of the pump. Clip valves and all other functions of the device are controlled by means of a computerized console. The use of clip valves reduces the cost of the disposable part of the device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.