Abstract

We have proposed an optical cavity based biosensor using a differential detection method for point-of-care diagnostics. For experimental demonstration of the proposed device using a two-laser system through refractive index measurements, an optical cavity structure is designed, fabricated, and measured. The differential value calculated with intensities of two different wavelengths has a larger responsivity with respect to the refractive index change as compared with the intesity of an individual wavelength. However, the repeatability test shows that the two-laser system has a tight fabrication tolerance due to its small dynamic range. In this paper, we introduce a three-laser system for an optical cavity sensor with a chained differential detection method and present the experimental measurement results showing a large dynamic range and large fabrication tolerance. The measured dynamic range of this system is more than three times greater than that of the two-laser system. The repeatability test using the three-laser system demonstrates a larger fabrication tolerance as a result of the increased dynamic range.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.