Acrylamide is a neurotoxin that is potentially carcinogenic to animals. In processed food, it is a side product of the Maillard reaction that is primarily formed when a carbohydrate-reach food is fried or baked. In this work, a sensitive and low-cost biosensor was prepared on a simple carbon paste electrode. Hemoglobin was initially fixed on a hybrid nanocomposite composed of iron magnetic nanoparticles with a layer of chitosan. The nanocomposite/hemoglobin was mixed with carbon paste to produce a biosensor. The device presented repeatability for both measurement and construction, with relative errors of 4.14% and 4.68%, respectively. The calibration curve presented linearity (R2 equal to 0.9945) for acrylamide concentrations ranging from 10 to 171 nmol L−1. The limits of detection was 0.06 nmol L−1, respectively, with analytical recoveries from 90.83% to 103.64%. The device was successfully used for the analysis of French fries . These results show that a simple and sensitive device was constructed, which is suitable to monitor the levels of acrylamide in food.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.