The use of online mass spectrometry for detecting volatile organic compounds (VOCs) has proven to be a powerful technique, allowing for real-time analysis of many chemical and biochemical processes. Unfortunately, online mass spectrometry has had limited application due to high instrument costs and limited availability. Here, we detail the design, construction, and performance characteristics of a custom ion-molecule reactor retrofitted to a commonly used single quadrupole mass spectrometer to operate as an online chemical ionization mass spectrometer (CIMS). This low-cost modified CIMS is capable of limits of detection below 10 parts per trillion for select VOCs including dimethyl sulfide, dimethylamine, and trimethylamine.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.