Abstract

The UK contribution to achieving the ultra-low background conditions required inside the detectors of the SuperNEMO experiment are described. A dedicated facility has been established for the screening and selection of materials through gamma ray spectroscopy using germanium detectors. Initial results from two detectors are shown. The radon level inside the SuperNEMO detector must be less than 150 μBq/m3 in order to achieve the target sensitivity. A Radon Concentration Line (RnCL) has been developed capable of measuring radon levels in large gas volumes down to 5 μBq/m3, improving on standard state-of-the-art radon detectors by 3 orders of magnitude. The development, commissioning and first measurements of radon content using the RnCL are also presented.