Abstract

National Radiation Protection Institute in Prague is equipped with 14 HPGe detectors with relative efficiency up to 150%. Steel shielding with one of these detectors (relative efficiency 100%) was chosen to be rebuilt to decrease minimum detectable activity (MDA). Additional lead and copper shielding was built up inside the original steel shielding to reduce the volume of the inner space and filled with nitrogen by means of evaporating liquid nitrogen. MDA values decreased for Compton background up to 0.67 of original value.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.