Abstract
Radius and ulna fractures in two dogs were repaired by application of a bone plate to the cranial surface of the radius. In both dogs, two screws were placed through the radius and into the ulna. Return of function and bone healing occurred as expected; however, return of lameness at 5 and 7 months postoperatively was attributed to loosening of the transosseous screws. The lameness resolved following implant removal. Loosening of only the transosseous screws could have resulted from the normal motion between the radius and ulna.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.