Abstract
In this paper, a novel approach based on frequency upconversion in ultra-thin nonlinear crystals is investigated for use in high-resolution infrared (IR) microscopy in the 5–12 µm range, an important domain for biomedical research. Traditional IR imaging encounters spatial resolution constraints due to diffraction, which are addressed via upconversion imaging using ultra-thin crystals. The present work combines a tunable IR quantum cascade laser and a short wavelength mixing laser to circumvent the classical resolution limit dictated by the Rayleigh criterion. A detailed numerical model for small signal upconversion imaging at μm-scale resolution shows good agreement with experimental data. The presented approach opens new avenues for IR applications for label-free biomedical diagnostics and spectral imaging.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.