Abstract

This article describes the long-term evaluation of a nanoceria-polyacrylonitrile (CeO2-PAN) composite sorbent-based (68)Ge/(68)Ga generator reported. This generator used the new CeO2-PAN composite sorbent for preparation of the (68)Ge/(68)Ga generator. Since this sorbent has not been previously evaluated, a thorough long-term evaluation of the performance of the generator is necessary to ensure its applicability for clinical practice. The performance of the generator was evaluated in terms of (68)Ga yield, (68)Ge breakthrough, radioactive concentration of the (68)Ga solution, and suitability of the (68)Ga for the preparation of (68)Ga-labeled tracers. The (68)Ge/(68)Ga generator was able to provide a (68)Ga activity with consistent yields (>70%) and having acceptable radionuclidic (<10(-4)% of (68)Ge breakthrough), radiochemical, and chemical purities for an extended period of time. The eluted (68)GaCl3 is useful for the majority of the (68)Ga complexation chemistry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.