Abstract

We analyzed serial data in patients with clinically stable monoclonal gammopathy to determine the total variation of serum M-spikes [measured with serum protein electrophoresis (SPEP)], urine M-spikes [measured with urine protein electrophoresis (UPEP)], and monoclonal serum free light chain (FLC) concentrations measured with immunoassay. Patients to be studied were identified by (a) no treatment during the study interval, (b) no change in diagnosis and <5 g/L change in serum M-spike over the course of observation; (c) performance of all 3 tests (SPEP, UPEP, FLC immunoassay) in at least 3 serial samples that were obtained 9 months to 5 years apart; (d) serum M-spike ≥10 g/L, urine M-spike ≥200 mg/24 h, or clonal FLC ≥100 mg/L. The total CV was calculated for each method. Among the cohort of 158 patients, 90 had measurable serum M-spikes, 25 had urine M-spikes, and 52 had measurable serum FLC abnormalities. The CVs were calculated for serial SPEP M-spikes (8.1%), UPEP M-spikes (35.8%), and serum FLC concentrations (28.4%). Combining these CVs and the interassay analytical CVs, we calculated the biological CV for the serum M-spike (7.8%), urine M-spike (35.5%), and serum FLC concentration (27.8%). The variations in urine M-spike and serum FLC measurements during patient monitoring are similar and are larger than those for serum M-spikes. In addition, in this group of stable patients, a measurable serum FLC concentration was available twice as often as a measurable urine M-spike.