Abstract

We propose an absolute distance measurement method using alternately oscillating optoelectronic oscillators (OEOs) with high speed, high precision, and long range, and describe the dynamic characteristics of the measurement system. Measurement and reference OEOs are oscillated using a 2×2 optical switch, and rough and fine measurements are achieved by low- and high-order-mode oscillation. The distance is determined by the loop length difference between the two OEOs. OEO length control is not necessary, so the system is simple and the time per measurement is only 40 ms. The maximum measurement error is 3.4 µm with an emulated distance of 7.5 km, and the relative measurement accuracy reaches 4.5×10-10.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call