A 79-year old man with a pulsatile mass in his left groin was admitted to our institution in 2012. He had undergone an aortobifemoral bypass using a bifurcated Cooley double velour knitted Dacron graft for aortoiliac occlusive disease in 1982. Computed tomography detected a 34 mm para-anastomotic pseudoaneurysm in the left limb of the graft. Graft replacement was performed to prevent the aneurysm from rupturing, and we found that the old Dacron graft had ruptured longitudinally along its guideline. The disrupted portion of the old Dacron graft was excised and examined by scanning electron microscopy, which revealed the deterioration of the polyester filaments around the guideline.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.