Abstract

Intraoperative femoral fracture is a well-recognized technical complication of cementless total hip arthroplasty. This study was designed to establish an in vitro model for initiation and propagation of fractures of the proximal femur in cementless THA and to assess the effect of fracture fixation in the form of cerclage wiring and drilling a hole in the distal extent of the fracture line. Fourteen human anatomic femur specimens were studied. Longitudinal cracks were made and propagation was performed on a materials testing system machine. A drill hole at the tip of a longitudinal crack does not prevent crack propagation. However, cerclage wiring has a statistically significant effect (p less than 0.025) on the ability of the fractured femur to withstand increased load.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call