Abstract

We reviewed charts and radiologic studies of 30 patients operated upon by ADR with Mobi-C® in single level since 2006. All patients had healthy cervical facet joints (less than or equal to grade 1 according to grading systems for cervical facet joint degeneration) preoperatively. We assessed clinical outcomes with NDI and VAS on neck and arm over follow-up and also measured ROM at implanted segment on dynamic radiographs during follow-up. The mean follow-up period was 42.4 ± 15.9 months. We then assessed the linearity of changes in ROM at implanted segment through linear mixed model. All patients showed significantly improved clinical outcomes. ROMs at implanted segment were maintained at slightly increased levels until 24 months postoperatively (P = 0.529). However, after 24 months, ROMs at implanted segment decreased significantly until last follow-up (P = 0.001). In addition, the decreasing pattern after 24 months showed a regular regression (P = 0.001). This decline was correlated with decline of extension angle at implanted segment. Based on this regular regression, we estimated that ROMs at implanted segments would be less than 2 degrees at 10.24 years postoperatively. Even though implanted segment maintains its motion for some length of time, we could assume that an artificial disc would have limited life expectancy correlated with the decline of extension angle.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.