Abstract

Dual mobility THA implantations are increasing. THA survival of more than 15years is considered long-term survival based on the Charnley prosthesis, which has a median wear rate of 39 to 98mm3/year. What is the median wear rate of dual mobility THA cups and its repartition on the liner volume? What are the factors influencing dual mobility wear? Thirty-five explants were analysed. The liners were entirely mapped with a fully automatic stripe light scanner that uses green LED light. The liners had a median wear rate of 38mm3/year. The distribution of the wear between the two bearings varied greatly. No factor was found influencing wear volumes. The median annual wear of dual mobility liners is at least as low as the wear of cemented polyethylene liners and lower than equivalent cementless liners. The two articulations of the dual mobility THA do not cause more wear. Level IV - case series.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.