Abstract
Testing computer-controlled linear accelerators for patient safety and proper patient dose delivery requires that certain beam characteristics be monitored over an extended period of time. Computer-controlled conformal radiation therapy using asymmetric collimator jaw settings necessitates stable symmetric treatment beams. Long term beam symmetry measurements have been performed on a Philips SL20 dual energy computer-controlled linear accelerator. Symmetry in both the radial and transverse axis of each x-ray beam was monitored for eight gantry positions. These measurements were undertaken to determine the effectiveness of the SL20 beam steering system during dose delivery of 50 monitor units (MU) per field. Evaluation of the data shows that careful beam steering setup procedures result in x-ray beams in which fluctuations in symmetry as a function of gantry angle are within +/- 1.5%. Day to day instabilities produce a total overall variation in beam symmetry on the order of +/- 2.0%. Results suggest the measurement of symmetry as a function of gantry position become a routine quality assurance procedure for this accelerator.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.