Abstract
Transvene models 6936/6966, a coaxial polyurethane ICD lead, may be prone to structural failure. These models comprise 54% of ICD lead failures in the authors' Multicenter Registry database. Because ICD leads perform a vital function, the clinical features, causes, and probability of Transvene 6936/6966 lead failure were determined. The Registry and United States Food and Drug Administration databases were queried for the clinical features and structural causes of the Transvene 6936/6966 lead failure, and a five-center substudy estimated the survival probability for 521 Transvene 6936/6966 implants. The mean time to failure was 4.8 +/- 2.1 years, and the estimated survival at 60 and 84 months after implant were 92% and 84%, respectively. Oversensing was the most common sign of failure (76%), and 24 patients experienced inappropriate shocks. The manufacturer's reports indicated that high voltage coil fracture and 80A polyurethane defects were the predominant causes of lead failure. Transvene models 6936 and 6966 coaxial polyurethane ICD leads are prone to failure over time. Patients who have these leads should be evaluated frequently. Additional studies are needed to identify safe management strategies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.