Abstract

We evaluated the outcomes of total hip arthroplasty in 67 patients (86 hips) with congenital hip disease and excessive abnormal anatomy of the proximal femur with the use of custom-made femoral stems. The design of the stem was based on CT data following the principles of CAD-CAE-CAM technique. No serious complications attributed to the femoral stem were seen. Within a median follow-up of 127.5months the 10-year survival of any of the components was 95.4% and respective value when aseptic loosening of the stem was considered was 98.1%. Patients with high dislocations had a 10-fold risk for loosening compared to those with low dislocations. No other parameter was associated with outcomes. The clinical and radiological evaluation was in consistency with the above outcomes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.