Abstract

Carotid angioplasty with stent placement has been proposed as an alternative method for revascularization of carotid artery stenosis. A novel stent with a laser-cut, rolled sheet of Nitinol (EndoTex Interventional Systems, Inc., Cupertino, CA) has been developed to customize treatment of stenotic lesions in carotid arteries utilizing a single stent, designed to adapt to multiple diameters and to tapered or nontapered configurations. The purpose of this study is to evaluate the conformability and vascular response to a novel stent in a chronic porcine carotid model using serial three-dimensional intravascular ultrasound (IVUS) analysis as well as histological examination. Ten Yucatan pigs underwent stent implantation in both normal carotid arteries with adjunctive balloon angioplasty. Three-dimensional IVUS analysis was performed before stent implantation, after adjunctive balloon angioplasty, and at follow-up [1 month (n = 6), 3 months (n = 6), or 6 months (n = 8)]. Histological examination (injury score, percent plaque obstruction, and qualitative analysis) was also performed. All stents were successfully deployed and well apposed in different sized vessels (lumen area range: 19-30 mm(2)). Volumetric IVUS analysis showed no significant difference between the lumen areas before stent implantation and after adjunctive balloon angioplasty and no stent area change at each follow-up point compared to immediately postprocedure. Histological examination revealed minimal injury and neointimal hyperplasia at each follow-up point. In the chronic porcine carotid model, the novel stent system demonstrated good conformability, resulting in minimal vessel injury and neointimal formation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call