Abstract

BackgroundThe newly-designed alumina ceramic medial pivot total knee prosthesis was introduced to reduce polyethylene wear and better fit the anatomical morphology of the Asian population. This study aimed to clarify the long-term clinical results of alumina medial pivot total knee arthroplasty over a minimum follow-up period of 10 years.MethodsThe data of 135 consecutive patients who underwent primary alumina medial pivot total knee arthroplasty were analyzed in this retrospective cohort study. Patients were examined over a minimum 10-year follow-up period. The knee range of motion, Knee Society Score (KSS) knee score, Knee Society Score function score, and radiological parameters were assessed. The survival rate was also evaluated by using reoperation and revision as endpoints.ResultsThe mean follow-up period lasted 11.8 ± 1.4 years. Patients who were not followed accounted for 7.4% of the total cohort. Knee and function scores of KSS improved significantly following total knee arthroplasty (P < 0.001). In 27 individuals (28.1%), a radiolucent line was observed. Aseptic loosening occurred in three cases (3.1%). The survival rates for reoperation and revision were 94.8% and 95.8% 10 years after the operation, respectively.ConclusionsDuring a minimum 10-year follow-up period, the present model of alumina medial pivot total knee arthroplasty showed good clinical outcomes and survival rates.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call