Abstract

We are developing an axial flow blood pump with Nimbus Inc. (AxiPump). For in vivo evaluation the AxiPump has been used as a left ventricular assist device with a left ventricular and descending aorta cannulation and implantation in a small pocket on the left lateral abdominal wall just posterior to the costal margin. Electrical and flow probe leads exit the body transcutaneously. A purge line that delivers the purge fluid for lubrication of the seal between the rotor and stator bodies in the purge fluid bearing system is tunneled with the other leads. Following acute animal studies, 3 animals have been supported for over 1 month with this AxiPump system. All laboratory results were within normal limits except during a recovery period from surgical damage. Hemolysis was not a serious problem. In the first case, the purge system failed at 28 days, and in the second and third cases, the nonpurge bearing system worked well for 57 and 52 days, respectively. Bearings are still under development in this kind of pump. However, this success encourages us to improve the AxiPump as a long-term assist device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.