Abstract

Experience with Type C thermocouples operating for long periods in the 1400 to 1600 °C temperature range indicate that significant decalibration occurs, often leading to expensive downtime and material waste. As part of an effort to understand the mechanisms causing drift in these thermocouples, the Idaho National Laboratory conducted a long duration test at 1500 °C containing eight Type C thermocouples. As report in this document, results from this long duration test were adversely affected due to oxygen ingress. Nevertheless, results provide key insights about the impact of precipitate formation on thermoelectric response. Post-test examinations indicate that thermocouple signal was not adversely impacted by the precipitates detected after 1,000 hours of heating at 1,500 °C and suggest that the signal would not have been adversely impacted by these precipitates for longer durations (if oxygen ingress had not occurred in this test).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.