Abstract
This study investigated 2 methods of reference base fixation for computer navigation markers for computer-navigated total knee arthroplasty. Five cadaveric specimens were used to test a 1-pin and 3-pin base system. A navigation system (Stryker Navigation, Kalamazoo, Mich) was used for testing with applied loads and torques to the reference base. Changes in distance from a verification point as well change in alignment were recorded. The change in distance to a reference point as well as the change in alignment data was significantly different at 65 N of applied load with the 3-pin construct being more stable ( P = .02). The results suggest that 3-pin fixation in the metaphyseal portion of the distal femur is more stable than a single-pin bicortical construct.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.