Abstract

BackgroundThe purpose of this study was to determine whether intramedullary administration of extended-release minocycline microspheres would affect osseointegration. MethodsTwenty-two rats were randomized to minocycline or saline femoral intramedullary injection followed by implantation of titanium alloy rods. Following euthanasia at four-weeks, pushout testing was performed and bone-volume-fraction assessed. ResultsPushout strength was marginally greater in minocycline-treated implants (122.5 ± 39.1 N) compared to saline (96.9 ± 26.1 N) (P = 0.098). No difference was observed in energy to maximum load, mean stiffness, or peri-implant bone-volume-fraction (P > 0.05). ConclusionsPeri-implant minocycline administration did not impair implant fixation strength or peri-implant bone-volume, supporting its potential utility as an adjunct to intramedullary implants.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.