Abstract

In this work, we propose an intuitive and easily implementable approach to model and interpret scanning Kelvin probe microscopy images of insulating samples with localized charges. The method, based on the image charges method, has been validated by a systematic comparison of its predictions with experimental measurements performed on charge domains of different sizes, injected in polymethyl methacrylate discontinuous films. The agreement between predictions and experimental lateral profiles, as well as with spectroscopy tip–sample distance curves, supports its consistency. The proposed procedure allows obtaining quantitative information such as total charge and the size of a charge domain and allows estimating the most adequate measurement parameters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.