Abstract
A movable, actively decoupled surface coil has been employed to obtain a localized 1H NMR spectrum from the lumbosacral spinal cord of a live Lewis rat. A volume selective 'VOSY' normally spelled out as 'volume selective spectroscopy' spectroscopy pulse sequence that incorporates 'phase ramped' selective RF pulses, has been used to minimize random phase jitter in the NMR signal as a result of the large frequency shifts required to locate the voxel in the center of the cord while using intense gradient pulses. Spectra from 13-microliters voxels in healthy rats and in rats inoculated with guinea pig spinal cord and complete Freund's adjuvant, resulting in experimental autoimmune encephalomyelitis, are shown.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.