Abstract

Modern lung counter systems which use several HPGe detectors can be used to locate a “hot spot” of uranium in the lungs by the use of vector analysis. If there is only one hot particle this localization cancels the need to assume homogenous distribution of the contamination for the calculation of its activity and enables more accurate determination of the activity of the contamination. In previous paper it was shown that the use of eight-dimensioned vectors representing simultaneously the count rates of the peaks at the energies of 92 keV (due to 238U) and 186 keV (due to 235U) leads to better results than using only one of the peaks (four-component vectors). In the present work it is shown that the possible variability of the isotopic composition of the uranium source has a major effect on the process of localization of the “hot spot” in the lungs. A way to eliminate this effect is proposed by separate normalization of the count rates of each energy peak, but still using eight-dimensioned vectors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.