Abstract

In the community of localization-based super-resolution microscopy (or called localization microscopy), it is generally believed that the emission of single molecules is so weak that an EMCCD (electron multiplying charge coupled device) camera is necessary to be used as the detector by eliminating read noise. Here we evaluate the possibility of a new kind of low light detector, scientific complementary metal-oxide-semiconductor (sCMOS) camera in localization microscopy. We demonstrate experimentally that sCMOS is capable of imaging actin bundles with FWHM diameter of 37 nm, evidencing the capability of sCMOS in localization microscopy. We further characterize the noise performance of sCMOS and find out that, with the use of a bright fluorescence probe such as d2EosFP, localization microscopy imaging is now working in the shot noise limited region.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.