Abstract

A CPMG pulse sequence incorporated with a DANTE-Z slice selection scheme for measuring spatially-resolved T2 distributions has been presented. The DANTE-Z pulse train with sinc-modulated pulses selects a single, quasi-rectangular slice of less than 0.8cm wide at an arbitrary position over a 6-cm long sample. The measured T2 distributions are of almost the same quality as regular (bulk) CPMG measurements, with the lower T2 limit being as good as c.a. 0.5ms. The sequence can be found useful as a supplement or alternative to MRI-based techniques for T2 mapping in short relaxation time samples (water-saturated rocks, building materials, wood, food products, rubbers, etc.), particularly when T2 is required to be measured at only few positions along the sample and the resolution of ∼1cm is acceptable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.