Abstract

The aim of this work was to establish local diagnostic reference levels (DRLs) for retrograde wedge portography (RWP) performed on pediatric patients assessing the usefulness of radiation dose monitoring software in the establishing process. Between September 2016 and April 2020, 66 consecutive RWP were performed at a transplantation institute and were included in our study. Patients were divided in three groups according to age: n=25 infants, n=20 middle childhood and n=21 early adolescence. The third quartile of both Air Kerma at the reference point (Ka,r) and air kerma-area product (PKA) were evaluated to establish local DRLs (lDRLs). In addition, to control high Ka,r levels during procedures, the software notified to operators if Ka,r exceeded the dose 'alert' threshold set at 2 Gy. lDRLs were established for all three groups using PKA and Ka,r: infant group: 5.6Gy.cm 2 and 0.034Gy; middle childhood: 6.4Gy.cm2 and 0.018Gy and early adolescence: 12.8Gy.cm2 and 0.059Gy. The dose threshold 'alert' was never encountered (alert quota: 0%). The dose monitoring system supports the feasibility of accurate and easier lDRLs' establishment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.