Abstract

The aim of this work was to propose a preliminary local diagnostic reference levels (DRL) for pediatric interventional cardiology (PIC) procedures in Argentina, for different ranges of age and weight. This work has been conducted in the framework of the "Optimization of Protection in Pediatric Interventional Radiology in Latin America and the Caribbean" (OPRIPALC) program coordinated by the World Health Organization and the Pan American Health Organization in cooperation with the International Atomic Energy Agency to ensuring that radiation exposures of pediatric patients are the minimum necessary during fluoroscopy-guided interventional procedures. The local DRL values presented in this paper by weight group and age group were 7.1 Gy·cm2 (<5 kg), 10.7 Gy·cm2 (5-15 kg), 18.0 Gy·cm2 (15-30 kg), 15.9 Gy·cm2 (30-50 kg), and 28.2 Gy·cm2 (50-80 kg) and 5.3 Gy·cm2 (<1), 11.2 Gy·cm2 (1 to 5<), 19.6 Gy·cm2 (5 to 10<), and 21.4 Gy·cm2 (10 to 16<), respectively. Our dose results are among the values found in other international studies; however, there is great potential for dose optimization.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.