Abstract

A set of national diagnostic reference levels (DRLs) was established in Malaysia for a range of breast thicknesses in 2013, but no updates for full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT). Due to the increasing number of DBTs used and concern over radiation exposure, this study aimed to explore and establish local diagnostic reference levels for FFDM and DBT in Malaysia health facilities at different compressed breast thickness (CBT) ranges. The CBT, kilovoltage peak (kVp), Entrance surface dose (ESD), and average glandular dose (AGD) were retrospectively extracted from the mammography Digital Imaging and Communications in Medicine (DICOM) header. The 75th and 95th percentile values were obtained for the AGD distribution of each mammography projection for three sets of CBT range. The difference in AGD values between FFDM and DBT at three CBT ranges was determined. The DRLs for FFDM were 1.13 mGy, 1.52 mGy, and 2.87 mGy, while DBT were 1.18 mGy, 1.88 mGy, and 2.78 mGy at CBT ranges of 20–39 mm, 40–59 mm, and 60–99 mm, respectively. The AGD of DBT was significantly higher than FFDM for both mammographic views (p < 0.005). All three CBT groups showed a significant difference in AGD values for FFDM and DBT (p < 0.005). The local DRLs from this study were lower than the national DRLs, with the AGD of FFDM significantly lower than DBT.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call