Circulators are often used in RF systems; however, their size, cost, and weight increase greatly at lower frequencies ( 19 dB for 20% of variation of the load impedance. The circuit was implemented and tested with an ultrasound probe in a water tank. The results show that the circuit was able to send and receive signals simultaneously from a single ultrasound transducer element.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.