Abstract

Reducing radiation dose while maintaining adequate image quality on conventional and digital radiographs requires optimal use of the unit. Additional filtration avoids unnecessary tissue exposure and improves photon transmission. Automatic exposure control may reduce exposure and dose. The volume of tissue imaged must be limited by the use of diaphragms and shutters or compression. Sensitive detectors with increased photon detection also contribute to reduce dose. Radiographic films combined to rare-earth screens also afford a good photon-conversion efficiency. Large area flat panel amorphous silicon x-ray sensors may also reduce dose up to 50% compared to films. Finally, calculation of the Kerma-area product independent of the source distance constitutes an important indicator of radiation dose.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.