Abstract

This work presents the implementation of an RGB-D camera as a surrogate signal for liver respiratory-induced motion estimation. This study aims to validate the feasibility of RGB-D cameras as a surrogate in a human subject experiment and to compare the performance of different correspondence models. The proposed approach uses an RGB-D camera to compute an abdominal surface reconstruction and estimate the liver respiratory-induced motion. Two sets of validation experiments were conducted, first, using a robotic liver phantom and, secondly, performing a clinical study with human subjects. In the clinical study, three correspondence models were created changing the conditions of the learning-based model. The motion model for the robotic liver phantom displayed an error below 3 mm with a coefficient of determination above 90% for the different directions of motion. The clinical study presented errors of 4.5, 2.5, and 2.9 mm for the three different motion models with a coefficient of determination above 80% for all three cases. RGB-D cameras are a promising method to accurately estimate the liver respiratory-induced motion. The internal motion can be estimated in a non-contact, noninvasive and flexible approach. Additionally, three training conditions for the correspondence model are studied to potentially mitigate intra- and inter-fraction motion.