Abstract

We report a case of an 89-year-old female with active extravasation and hemoperitoneum from a liver laceration demonstrated on multidetector computed tomography (CT), attributed to the use of an automated mechanical cardiopulmonary resuscitation (CPR) device. Although iatrogenic internal injuries related to manual CPR and CPR devices have previously been reported [1, 2], there has been no reported CT evidence of liver injury related to automated CPR devices to the authors' knowledge. Imaging findings of complications related to the use of automated CPR devices are important to recognize and also help explain the possible mechanisms of injury. Liver injuries with active bleeding following CPR may have devastating consequences related to hemodynamic instability and may have an increased incidence when CPR is performed using an automated chest compression device.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.