Abstract

Dynamic contrast-enhanced MRI (DCE-MRI) registration is a challenging task because of the effect of remarkable intensity changes caused by contrast agent injections. Unrealistic deformation usually occurs by using traditional intensity-based algorithms. To alleviate the effect of contrast agent on registration, we proposed a DCE-MRI registration strategy and investigated the registration performance on the clinical DCE-MRI time series of liver. We reconstructed the time-intensity curves of the contrast agent through sparse representation with a predefined dictionary whose columns were the time-intensity curves with high correlations with respect to a preselected contrast agent curve. After reshaping 1D-reconstructed contrast agent time-intensity curves into a 4D contrast agent time series, we aligned the original time series to the reconstructed contrast agent time series through traditional free-form deformation (FFD) registration scheme combined with a residual complexity (RC) similarity and an iterative registration strategy. This study included the DCE-MRI time series of 20 patients with liver cancer. Qualitatively, the time-cut images and subtraction images of different registration methods did not obviously differ. Quantitatively, the mean (standard deviation) of temporal intensity smoothness of all the patients achieved 54.910 (18.819), 54.609 (18.859), and 53.391 (19.031) in FFD RC, RDDR, Zhou etal.'s method and the proposed method, respectively. The mean (standard deviation) of changes in the lesion volume were 0.985 (0.041), 0.983 (0.041), 0.981 (0.046), and 0.989 (0.036) in FFD RC, RDDR, Zhou etal.'s method and the proposed method. Our proposed method would be an effective registration strategy for DCE-MRI time series, and its performance was comparable with that of three advanced registration methods.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call